CRY3261-S01 Measurement Microphone Set
The CRY3261-S01 is a 1/2-inch free-field prepolarized measurement microphone and preamplifier set widely used in acoustic testing environments such as anechoic chambers and quiet rooms.


Interested
For more information on CRY3261-S01 Measurement Microphone Set or to request a quote, contact our team now:



The unique design of CRY3261-S01 allows it to have an extremely low self-noise, making it highly suitable for detecting and measuring ultra-low noise levels. Typical applications include measuring the self-noise of products, computer fan noise, transformer current noise, and evaluating environmental noise.
Unsure what measurement microphone best meets your requirements? Use our handy measurement microphone selection tool to discover which measurement microphone set is the best fit for you.
Technical Specification |
---|
Field Type | Free-field |
Nominal Sensitivity | 354mV/Pa, -9±2dB re 1V/Pa |
Frequency Response | 3.15Hz to 20kHz±3dB |
Polarization Voltage | 0V |
Dynamic Range(re.20uPa) | 7.5dBA to 115dB, Typical>6.5dBA |
Operating Temperature | -20 to +60℃ (-4 to +140℉) |
Temperature Stability | 0.012dB/℃ (-10 to +50℃) 0.007dB/℉ (+14 to +122℉) |
Static Pressure Stability | -0.01dB/kPa |
Operating Humidity Range | 0 to 90%RH no condensation |
Humidity Stability | < 0.1dB (0 to 90%RH no condensation) |
Pressure Equalization Vent | Rear vented |
IEC 61094-4 Type | WS2F |
Output Impedance | < 20Ω |
Maximum Output Voltage | >5.0Vrms |
Current Source | IEPE (2mA to 10mA) |
Typical Operating Current | 4mA |
Interface Type | DB9 |